Estimation of absorbed dose to the kidneys in patients after treatment with 177Lu-octreotate: comparison between methods based on planar scintigraphy

Abstract

Background

Lu-[DOTA0, Tyr3]-octreotate (177Lu-octreotate) is used to treat neuroendocrine tumors with high somatostatin-receptor expression. 177Lu-octreotate is mainly excreted via the kidneys, but to some extent, accumulates in the kidney cortex due to, e.g., tubular reabsorption. Renal toxicity is one of the main limiting factors in 177Lu-octreotate treatment. Further knowledge of the biodistribution and dosimetry of 177Lu-octreotate in individual patients is needed. The aim of this study was to estimate the absorbed dose to the kidneys and compare the results obtained with planar imaging and different dosimetric methods: (1) conjugate-view (CV) method using patient-specific kidney sizes, (2) PA method, based on posterior images only, (3) CV method with reduced number of time points (CVreduced data), and (4) CV method using standard kidney sizes (CVstandard size).

Methods

Totally, 33 patients each received 3.4 to 8.2 GBq of 177Lu-octreotate up to five times, with infusion of lysine and arginine to block the renal uptake. Whole-body planar gamma camera images were acquired on days 0, 1, 2, and 7. The 177Lu concentration in the kidneys was determined by the CV method, and the absorbed dose was estimated with patient-specific organ sizes. Comparison to the CV method was made using posterior images only, together with the influence of the number of time points and with standard organ sizes.

Results

Large interindividual variations were found in the time-activity curve pattern and in the absorbed dose to the kidneys using the CV method: 0.33 to 2.4 Gy/GBq (mean =  0.80 Gy/GBq, SD = 0.30). In the individual patient, the mean deviation of all subsequent kidney doses compared to that of the first administration was 1% (SD = 19%) and 5% (SD = 23%) for the right and left kidneys, respectively. Excluding data for day 7 resulted in large variations in the absorbed dose.

Conclusion

Large interindividual variations in kidney dose were found, demonstrating the need for patient-specific dosimetry and treatment planning.

Background

The somatostatin analog 177Lu-[DOTA0, Tyr3-octreotate (177Lu-octreotate) has been used to treat patients with somatostatin-receptor expressing neuroendocrine tumors with promising clinical results [14]. About one-third of the patients showed complete, or partial, tumor remission and many patients with advanced tumor burden experienced symptomatic relief.

After infusion of 177Lu-octreotate, 177Lu is mainly excreted via the kidneys but also reabsorbed and accumulated in the kidney cortex [510]. The mechanisms behind this accumulation are not fully understood, but studies on rats indicate that endocytosis either via the megalin-cubulin complex, or via ligand-specific endocytosis, pinocytosis, organic anion and cation transporters families, and oligopeptide transporters, are involved [5, 8, 1013].

Renal toxicity is one of the main limiting factors in 177Lu-octreotate treatment [4, 14, 15]. To avoid adverse effects of 177Lu-octreotate and to determine the tolerance dose to kidneys, there is a need for better knowledge of the renal biodistribution, biokinetics, and dosimetry of 177Lu-octreotate and of relevant radiobiological and toxicity factors [4, 5, 11].

Detailed dosimetric calculations are time- and labor-intensive. The most commonly used methods include anterior and posterior planar scintigraphy (static images or whole body scans) and the conjugate-view method (CV method) [16]. Accurate determination of absorbed dose by the CV method requires scintigraphic data from multiple time points over a reasonably long time interval following administration of the radiopharmaceutical, accurate calibration of the gamma camera system, appropriate corrections for background activity, attenuation, and scatter, and reliable estimation of organ masses. If possible, less time-consuming routine methods should be defined with adequate accuracy in activity quantification and absorbed dose determination.

Dosimetric estimations for patients treated with 177Lu-octreotate have previously been published [15, 1720]. Results from these publications demonstrate variations in absorbed dose to the kidneys between clinics and between methods used within a clinic, with in general lower values obtained from SPECT images compared to planar images. Differences were also obtained using different methods based on planar images and the CV method. Since the latter method is most widely used, it is of interest to study how the choice of parameters influences the results, and also to study if simplifications can be applied to reduce the calculation time.

The aim of this study was to determine the mean absorbed dose to the kidneys in 33 patients with neuroendocrine tumors treated with 177Lu-octreotate using the CV method. The results were compared between patients and between treatment cycles given to the same patient. Furthermore, results from (1) the CV method (with patient specific organ sizes) were compared with those from (2) the PA method (a simplified method, based on posterior images only), (3) the CV method with reduced number of time- points (CVreduced data) (one time point fewer), and (4) the CV method using standard organ sizes (CVstandard size) (to study the importance of patient-specific organ sizes).

Methods

Patients

A total of 33 consecutive patients with neuroendocrine tumors were included, 20 men and 13 women, aged 40 to 80 years (mean 63 years). The lesions included endocrine pancreatic tumors, gastrointestinal carcinoid tumors, and one lung carcinoid tumor (Table 1). Patients recruited had adequate bone marrow, kidney (GFR > 40 mL/min), and liver functions, and high somatostatin receptor expression by the tumor tissue, demonstrated by high tumor uptake of 111In-octreotide (OctreoScanTM, Covidien, St. Louis, USA) vs. normal tissues such as kidney and liver based on a diagnostic scintigraphic examination. The estimated 177Lu concentration in tumor tissue always exceeded that in the liver. Some of the patients had previously undergone hepatic artery embolization, and 11 patients were treated by chemotherapy, which was terminated at least 3 months before 177Lu-octreotate treatment. One patient (no. 5) had only one kidney. License for the treatment with 177Lu-octreotate was received for each patient individually by the Swedish National Medical Products Agency.

Table 1 Patient data, including age, sex, type of tumor, location of metastases, and earlier treatments

Radiopharmaceuticals and administration

177Lu-chloride and [DOTA0, Tyr3]-octreotate were obtained from the Nuclear Research & Consultancy Group (Petten, the Netherlands, distributed by IDB Holland, the Netherlands). [DOTA0, Tyr3]-octreotate was labeled with 177Lu according to the instructions by the manufacturer. The fraction of peptide-bound 177Lu was more than 99%, as determined by instant thin-layer chromatography with 0.1-M sodium citrate as the mobile phase.

Each patient received 3.4 to 8.2 GBq 177Lu-octreotate up to five times (treatment cycles), i.v. infused during 30 min (200 mL/h). The number of patients that received 1, 2, 3, 4, and 5 treatment cycles were 3, 7, 10, 12, and 1, respectively, and the average activity given at cycle numbers 1 to 5 were 7.8, 7.2, 7.2, 7.2, and 7.8 GBq, respectively. To reduce kidney uptake, the patient received an i.v. infusion with lysine 2.5% and arginine 2.5% in NaCl solution (250 mL/h) over 4 h starting 30 min before the administration of 177Lu-octreotate. Before lysine-arginine infusion, the patient received 3 mg granisetron (Kytril®, Roche, Genentech, Inc., South San Francisco, CA, USA) i.v. against nausea. The total number of administrations and amount of 177Lu-octreotate at the last administration were adjusted not to exceed a total absorbed dose to the kidneys of 28 Gy. The 177Lu activity administered was determined using a carefully calibrated ionization chamber (Capintec CRC®-120, Scanflex Medical AB, NJ, USA). Residual activity in the tube system was corrected for.

Scintigraphy

Two gamma camera systems were used: Picker IRIX system (Marconi, Philips, Holland) and Millennium VG Hawkeye system (General Electric Medical Systems, Milwaukee, WI, USA), both equipped with medium energy parallel-hole collimators. Planar whole body images were usually acquired by the IRIX system, while SPECT/CT images were acquired with the Millennium system. Image processing and analysis were performed on a GENIE Xeleris workstation (General Electric Medical Systems, Milwaukee, WI, USA).

One hour, and 1, 2, and 7 days after onset of the 177Lu-octreotate infusion anterior and posterior whole body scintigraphy was performed (scanning speed of 10 cm/min, 20% energy-window centered over the 208-keV photon peak, 256 × 1,024 matrix, pixel size of 2.06 mm (IRIX) and 2.21 mm (Millennium)). SPECT/CT was performed at day one (step-and-shoot acquisition mode: 128 × 128 matrix, 4.42-mm pixel size, 4.42-mm slice thickness, 180° gantry rotation, 3°/step, 30-s frame time duration, energy window setting as above). CT images were acquired using 140-kV tube voltage, 2.5 mAs, 2.6 rpm rotation speed, 256 × 256 matrix, 2.21-mm pixel size, and a 4.42-mm slice thickness. CT images were converted to 128 × 128 matrix and fused with SPECT images. The CT images were only used for the determination of the body thickness and thickness and volume of the kidney, and the SPECT images only for the localization of regions with specific or non-specific 177Lu uptake, overlying or present close to the kidneys.

Calibration of the gamma cameras

The sensitivity, k, and the effective linear attenuation coefficient, μ, for 177Lu were determined for the two gamma cameras [16]. A 177Lu source (25 ml, 330 MBq) was positioned at 0.5 to 19-cm depth in a polystyrene phantom 1.05 g/cm3. Static images were acquired for 300 s, using the same matrix, pixel size and energy settings as for the whole body images. The number of counts in a region-of-interest (ROI) around the source in the images was determined and plotted versus the depth. An exponential curve was fitted to the data points; k was determined as the extrapolated intersection with y-axis and μ was obtained from the exponential constant of the fitted curve [16, 21]. For the IRIX and Millennium systems, k was 10.9 cps/MBq and 7.66 cps/MBq, and μ was 0.119 and 0.118 cm−1, respectively.

Dosimetry

For each treatment cycle, ROIs were drawn around each kidney on the posterior images from the time point when the kidneys were best visualized (day 1 or 2). These ROIs were then used for all posterior images acquired at the other time points and then mirrored and used for anterior images. In all but one patient, high 177Lu activity in the liver, spleen, intestines, and/or tumor tissue was overlapping one or both kidneys based on SPECT images and visual analysis of planar images. Then, small ROIs were also drawn over part of the kidney without overlap, and the number of counts was extrapolated to include the entire kidney, assuming similar counts per pixel.

For background subtraction, a small circular ROI was placed adjacent to each kidney. The mean counts/pixel was determined and scaled to a body thickness excluding the kidney thickness (using the CT image, and assuming homogeneous background activity concentration within the volume of interest defined by the background ROI) and then subtracted from the mean counts/pixel in the kidney ROI [21, 22]. The 177Lu activity in each kidney was determined in four ways: (1) by the CV method using anterior and posterior data with patient-specific organ sizes [16], (2) using only the posterior data (PA method) with patient-specific organ sizes, (3) by the CVreduced data method with patient-specific organ sizes with reduced number of time points (one less than in method (1)), and (4) by the CVstandard size method estimated with standard organ sizes.

For each time point, the activity in the kidney, A, was determined according to the CV method:

A = R A R P e μ T μ x k e μx / 2 e μ x / 2 ,
(1)

where R A and R P are the background-corrected (usually extrapolated, described above) number of counts in the kidney ROI from the anterior and posterior images, respectively. T is the anterior-posterior body thickness at the kidney level, and x is the anterior-posterior kidney thickness. k is the gamma-camera sensitivity, and μ is the mean linear attenuation coefficient for 177Lu.

The activity in the kidney using PA method:

A = R P e μ a μ x k 1 e μ x ,
(2)

where a is the distance from the back of the patient to the posterior edge of the kidney.

A mono-exponential function was fitted to the measured, decay corrected time-activity data for each kidney, with a mean fit of R2 of 0.94 ± 0.13 (SD). Two patients (nos. 15 and 27) had a somewhat poorer fit R2 < 0.5. The cumulated activity, the number of disintegrations from the time of start of infusion (t = 0) to infinity, , was then calculated

A ˜ kidney = 0 A 0 e λ eff · t d t = A 0 λ eff
(3)

where A 0 is the activity at time t = 0, and λ eff is the effective decay constant for the kidney.

The volume of each kidney was estimated from the CT images by approximation by an ellipsoid, giving the length, width, and thickness of the kidney. The mass was calculated assuming the tissue density 1.05 g/cm3 [23].

The mean absorbed dose for each kidney was determined

D ¯ kidney = A ˜ kidney n E ϕ m kidney ,
(4)

where nE is the mean energy emitted by electrons per decay (147 keV for 177Lu) [2426], ϕ is the absorbed fraction (assumed to be 1 for the electrons) [2729], and m kidney is the mass of the kidney.

To study the necessity of data points-in-time, the mean absorbed dose for each kidney per administered activity (D/A administered) was determined when omitting one of the data points (1 h, 1 day, 2 days, or 7 days) and compared with D/A administered based on data from all time points. In this sub-study, data from 9 treatment cycles (in eight patients) were not included, since the patients were not examined at all time points at those treatment cycles due to their medical condition.

For comparison between dosimetric estimations using patient-specific kidney size (mass and thickness) and standard kidney size, 10 of the patients (four female and six male patients) with different gender and kidney sizes were chosen, including patients with the smallest and largest kidneys. D/A administered was determined using the CV method with standard masses of the kidneys (310 g for men and 275 g for women, both kidneys, taken from ICRP 89) and body sizes (20 cm trunk and 6 cm kidney thickness for men; 18 and 5 cm, respectively, for women, adapted from MIRD Pamphlet No. 5 [23, 30]).

The contribution from activity in the remainder of the body to the absorbed dose to the kidney was estimated using the whole-body cumulated activity (assumed to be uniformly distributed and excluding the renal cumulated activity) and whole-body-to-kidney S values for 177Lu from RADAR/OLINDA [31].

Statistical analysis

Statistical analyses and linear regression analyses were performed using SPSS PASW Statistics 18 (SPSS Inc, Chicago, IL, USA). Differences between groups were analyzed using Student's t test; p values < 0.05 were considered statistically significant.

Results

When estimating the 177Lu activity in the kidneys from the planar images, regions with high 177Lu activity within organs and tissues overlapping the kidney ROIs (mainly the liver, spleen, intestines, and paraortic lymph node metastases) were excluded. Based on all total kidney ROIs defined, the area of ‘overlapping tissues’ varied from 0% to 90% of a ROI, with typical values of 50% to 70% (in 65% of the kidney ROIs).

Time-activity curves

There were differences in the shapes of the time-activity curves of the kidneys for different patients, also among different treatment cycles in the same patient (Figure 1). There were two main patterns, either with a maximum value at the first time point studied (1 h) or at the second or third time points (day 1 or day 2). The figure also demonstrates the differences between right and left kidneys and differences among treatment cycles (Figure 1A). In two of the patients (nos. 11 and 20), the uptake phase differed between the right and left kidneys.

Figure 1
figure 1

The177 Lu activity in the right and left kidneys given as percent of injected activity. The 177Lu activity in the right and left kidneys given as percent of injected activity (%IA) for (A) and (B) patient no. 8 (left and right kidneys, treatment cycles 1 to 4) and (C) for patient no. 24 (left kidney, treatment cycles 1 to 4).

Mean absorbed dose per unit administered activity

The mean absorbed dose per unit administered activity to the right and left kidneys, respectively, for all 33 patients included in the study are shown in Table 2 and Figure 2. The results show a large individual variation, with absorbed dose per unit administered activity of 0.33 to 2.4 Gy/GBq per treatment cycle for one kidney (mean value 0.80 Gy/GBq; SEM = 0.02, SD = 0.30). Considering treatment cycle 1 only, absorbed dose per unit administered activity was 0.38 to 1.7 Gy/GBq (mean value 0.84 Gy/GBq; SEM = 0.04, SD = 0.29). Patient no. 5 (with only one kidney) did not differ considerably in absorbed dose per unit administered activity to the single kidney (mean of three treatment cycles 0.69 Gy/GBq (SD = 0.11)) compared to that in patients with two kidneys.

Table 2 The mean absorbed dose per administered activity
Figure 2
figure 2

The mean absorbed dose per unit administered activity. The mean absorbed dose per unit administered activity, to the right and left kidneys, respectively, for all 33 patients included in the study. Results are given as mean values of all treatment cycles. Patient nos. 6, 15, and 33 only received one treatment cycle, and for patient no. 32 data for the right kidney could only be obtained from one treatment cycle. Patient no. 5 had only one kidney. For patient nos. 1, 4, 13, 14, 19, 23, 24, 29, 30, and 31 only one kidney dose could be evaluated due to high uptake in overlapping and adjacent tissues (see text). Error bars indicate ± 1 SD.

Differences in mean absorbed dose to kidney for each treatment cycle

D/A administered obtained for the right and left kidneys is compared in Figure 3 for each patient and treatment cycle. The mean deviation of all subsequent kidney doses from the kidney dose obtained from the first treatment cycle was 1% (range −31% to +47%) for the right kidney and 5% (range −26% to +93%) for the left kidney (Figure 3C,D). The results show no trend towards higher or lower values with increased treatment cycle number for the group, but rather a clear difference among patients.

Figure 3
figure 3

The mean absorbed dose per unit administered activity for each treatment cycle in each patient. The mean absorbed dose per unit administered activity given to (A) right kidney and (B) left kidney for each treatment cycle in each patient. The mean absorbed dose per unit administered activity normalized to the value for the first treatment cycle given to (C) right kidney and (D) left kidney for each patient. Each line corresponds to one patient and the different colors are used to help the reader to follow at least some of the lines to better see different patterns.

Comparison between the right and left kidneys

D/A administered for the right and left kidneys was compared (Figure 4). No statistically significant differences were found for the first treatment cycle (p = 0.43) or for all treatment cycles (p = 0.83) using Student's t test (paired, two sided).

Figure 4
figure 4

Comparison between the mean absorbed dose per unit administered activity to the left and right kidneys. Comparison between the mean absorbed dose per unit administered activity to the left and right kidneys for (A) all treatment cycles and (B) only the first treatment cycle. (C) Ratio between the absorbed dose to the right and left kidneys for all treatment cycles in the studied 22 patients.

Comparison between CV and PA methods

D/A administered was 0.60 Gy/GBq (SD = 0.17; range 0.29 to 1.2) using the PA method and 0.80 Gy/GBq (SD = 0.30; range 0.33 to 2.4) using the CV method. The correlation between the two methods was statistically significant: p < 0.001 for both the right and left kidneys using Student's t test (paired, two sided) (Figure 5).

Figure 5
figure 5

Correlation between PA and CV methods. Correlation between the mean absorbed dose per unit administered activity, estimated with the PA and the CV methods for the (A) right and (B) left kidneys.

Importance of data points-in-time

The uncertainty when omitting one of the data points (1 h, 1 days, 2 days, or 7 days) was initially studied on 163 dose estimations in 32 patients, when the absorbed dose could be determined from all points-in-time. Of those 163 dose estimations, 23 were excluded, since the fitted mono-exponential curve to the time-activity curve had a positive coefficient, because day 2 had a higher uptake compared to 1 h and 1 day, resulting in infinite cumulated activity when omitting data from 7 days. The results clearly show that omitting data from 7 days gave the largest deviation in D/A administered compared to D/A administered from all time points, with a mean deviation by a factor of 1.1 (SD = 0.34) and a maximum deviation by a factor of 2.3 (Figure 6). It should be noted that these values when omitting results from day 7 are underestimated due to the exclusion of data with D/A administered = .

Figure 6
figure 6

Mean absorbed dose per unit administered activity, excluding one of the data points (CV reduced data method). Mean absorbed dose per unit administered activity to the kidney determined excluding one of the data points (1 h, or 1, 2, or 7 days after administration) divided by the mean absorbed dose per unit administered activity including all data points. Error bars indicate ± 1 SD. Note that the value, when excluding the day 7, data point is underestimated since values of infinity were excluded.

Comparison with fixed kidney mass and body size

The estimated kidney mass from CT images for the selected 10 patients was 88 to 234 g, the estimated trunk thickness was 16.7 to 26.8 cm for men and 16.4 to 21.4 for women, and the kidney thickness was 5.2 to 6.8 cm for men and 4.3 to 6.6 for women, used in the CVstandard size method with patient specific organ sizes (Table 2). Using the CVstandard size method with standard organ sizes, D/A administered was underestimated by about 50% for patients with the smallest kidneys (<100 g) and overestimated by about 50% for patients with the largest kidneys (>150 g) compared to D/Aadministered for the CV method with patient-specific dimensions.

Contribution from 177Lu distributed in the remainder of the body

The contribution from electrons and photons from the remaining tissues (total body except the kidneys) was found to be about 2% of absorbed dose per unit administered activity to kidneys.

Discussion

The present study focused on D/A administered to kidneys in patients with neuroendocrine tumors after fractionated treatment with 177Lu-octreotate. The 177Lu activity was determined using planar whole body scintigraphy and the CV method. D/A administered in our study (0.80 Gy/GBq) was similar to the results from other studies (0.65 to 1.15 Gy/GBq) (Table 3) [3, 15, 1720, 32].

Table 3 Summary of data obtained from different studies on kidney dosimetry for 177 Lu octreotate

When comparing the results from different studies, differences in kidney protection method, administered activity per treatment cycle, imaging technique, and methods for image analysis and dosimetry should be considered (Table 3). In one of the studies, different methods for imaging and dosimetric analysis were compared on the same patients, thus keeping the biological parameters constant: the lowest D/A administered was obtained using SPECT (0.81 to 0.90 Gy/GBq) compared to planar imaging and CV method (0.97 to 1.2 Gy/GBq) [20]. Similar results were obtained in another study despite the large uncertainty associated with the planar imaging-derived dose estimates in that study [19]. Furthermore, the importance of using an optimal background correction method was demonstrated [20].

One drawback using planar images and CV method is difficulties with high and unpredictable time-dependent activity uptake in overlapping normal tissues and tumors. We tried to avoid the problem using small ROIs over parts of the kidney with no overlap, assuming homogeneous activity distribution within each kidney. We found that only one kidney could be evaluated using the whole kidney area. Due to high overlapping uptake making it impossible to outline the kidneys in the images, 34 dose estimations could not be done (eight patients for the right kidney: patient nos. 4, 14, 19, 23, 24, 29, 31, and 32 (treatment cycle 2), and three patients for the left kidney: patient nos. 1, 13, and 30). This problem would be reduced using SPECT.

D/A administered varied between 0.33 and 2.4 Gy/GBq in the present study using the same method (CV method), which clearly underlines the value of individual dosimetry. We found that time-activity curve forms can differ considerably among patients and between different treatment cycles in the same patient. A similar study on 61 patients followed up to 68 h showed that the 177Lu activity in kidneys had a fast followed by slower clearance, with a maximal uptake of 4.0%IA [17]. This time course was seen in 21 of our 33 patients; for 12, patients we found a slower uptake phase during the first 1 to 2 days after 177Lu-octreotate administration. Similar differences between patients have previously been demonstrated for 111In-octreotide [33]. Furthermore, in two patients, we observed distinctly different time-activity curves for the right and left kidneys, maybe due to differences in kidney physiology or differences in the time-dependent activity in overlapping organs such as the intestines (see below). Ideally, multi-exponential curve fitting would have been preferable in the present study, at least for a few cases, but mono-exponential curve fitting has been routinely used in many studies and clinics, and worked well for most of the estimations in this application.

It is generally assumed that D/A administered to the right and left kidneys is the same within a patient, and this assumption is used when problems with radioactivity in tissues overlying one of the kidneys occur [17]. Although we found about 12% lower values for the left kidney than for the right kidney for the patient group, this difference was not statistically significant (p = 0.83, paired t test). Monte Carlo simulations on a phantom using the CV method demonstrated that the left kidney was less influenced by radioactivity in the intestines (especially colon) than the right kidney, and that the lowest uncertainty in activity determination in an organ without overlapping activity is ±10% if appropriate attenuation and scatter corrections are used [34]. The uncertainty in kidney dosimetry for 177Lu-octreotate would most probably be somewhat higher, and the side difference for most patients in the present study was within ±20%, which we think reflects the total uncertainty in the present study, but values up to 70% was found (Figure 4). Due to several reasons, we believe that the absorbed dose to the kidneys should, if possible, be determined separately for each kidney. There might be individual differences in surrounding/overlapping tissues with high 177Lu activity concentration for the right kidney, e.g., liver with metastases, and for the left kidney, the spleen, and for both kidneys, activity in the intestines and paraortal lymph node metastases. Another reason is that the kidneys might have different functions, leading to differences in uptake and retention.

When comparing D/A administered for each kidney and treatment cycle, we found no obvious change in the value with increased number of treatment cycles for the group of patients; for most of the patients, the differences were within the uncertainty (discussed above) related to 177Lu uptake and retention in surrounding (background) tissues. Others have found that the contribution from each treatment cycle to the total absorbed dose was uniform (difference within 10%), but for two patients, the contribution differed within 20% mainly due to differences in the kinetics [20]. In the present study, large individual variations were found for some patients (differing by up to 93% from the first treatment cycle), indicating and in some patients demonstrating a trend of increased, or decreased, kidney uptake and retention with number of treatment cycles (Figure 3). There are, however, possible sources of errors, or rather uncertainties, in the dose estimations between the treatment cycles that might overestimate the differences for a single patient: (1) The position of the ROI over the kidney might vary somewhat between subsequent imaging in a treatment cycles and/or between subsequent cycles, due to differences in patient positioning in the camera, different position and movement of the kidney (e.g., due to respiration). (2) The position of background ROIs, due to variations in radionuclide accumulation in regions surrounding the kidneys between subsequent image time points and/or treatment cycles. (3) There are uncertainties in measurement of residual activity in vials and infusion tubes and, hence, in the resulting activity administered. These results demonstrating differences in kidney dose for different cycles should be further examined in a future radiobiologically oriented study.

The results clearly demonstrate the importance of data from late time points. In cases when the maximum uptake value occurred at day 1 or 2, late time points are especially important to obtain a realistic extrapolation of the activity with time, in particular for, radionuclides with a relatively long physical half-life, such as 177Lu. Our finding demonstrating a large uncertainty when omitting data from day 7 (Figure 6) is also an underestimation since 23 kidney dose values were infinite and including them would give a mean value of infinity. The previous studies with no late images (only up to 48 or 68 h after infusion) reported similar D/A administered levels as studies with data collected at much later points-in-time (Table 3). The reason for this is not obvious to us, since we clearly noticed the importance of data from later points-in-time for correct dose estimation in several patients.

The determination of the activity concentration in the kidneys using the CV method is rather time-consuming, and a simpler, faster method is highly desirable. Therefore, we compared the D/A administered determined by the PA method with that based on the more correct CV method. We found a clear correlation between the values obtained with the two methods, but the linear regression coefficient was far from unity, demonstrating that the PA method gave too low D/A administered values. The PA method requires detailed knowledge of the depth and thickness of the kidneys and assumes homogeneous activity distribution together with accurate attenuation correction. Estimations based on five patients indicate a difference of about 0.12 Gy/GBq/cm of kidney thickness (data not shown). Since the performance of the PA method was poor, CV method is preferred because the method is independent of the depth of the kidneys, and that a correct attenuation correction needed for the PA method might be complicated and require Monte Carlo simulations.

When calculating the absorbed doses to kidneys with the CVstandard size method using standard kidney size (MIRD5 and ICRP 89) instead of patient-specific body and kidney sizes, large differences were found for patients with small or large kidney volumes. In agreement with a study on 86Y/90Y-DOTATOC, we thus found that individual kidney volume determination is needed for a more precise estimation of absorbed dose to the kidneys [14].

Conclusion

In conclusion, a large variation in absorbed dose per unit administered activity was found between patients (up to a factor of 8) and between administrations in the same patient (up to a factor of 2). The importance of patient-specific kidney and body sizes was demonstrated for kidney dose estimation of 177Lu-octreotate. There was a clear correlation between the absorbed dose determined by the simpler PA method and the more labor-demanding CV method, but the PA method gave too low values of the kidney dose. CV method is mathematically correct but assumes homogeneous activity concentration with depth within the kidney. One advantage with the CV method is that the AP position of the kidney does not influence the results. PA method is also mathematically correct and also assumes homogeneous activity concentration with depth within the kidney. The disadvantage with the method is that the AP position clearly influences the results and the AP position must be exactly defined. In most patients, the kidney axes are usually tilted compared to the back, which makes the determination of the AP position even more difficult. Therefore, the CV method is still preferred due to the higher accuracy expected. The need for late data points for better dose estimation for 177Lu-octreotate is also shown.

The results clearly demonstrate the importance of individualized dosimetric calculation for critical normal tissues for optimal treatment planning using 177Lu-octreotate. Underestimation of the actual absorbed dose to the kidneys may lead to renal failure, while overestimation of the absorbed dose would result in under-treatment of the tumor tissue and reduced treatment efficacy. To obtain accurate dosimetry, optimized activity quantification methods should be used, including methods used for scintigraphy (planar, SPECT), delineation of the organ, position of background ROI, attenuation and scatter correction, calibration, and late time points for imaging.

Further efforts should be made to define optimal but realistic dosimetric methods for the kidneys useful for routine treatment planning. The accuracy that is needed will depend on the knowledge of the tolerance dose or expected kidney toxicity at different absorbed dose (or BED) levels in the individual patient, knowledge that is still limited but needs to be obtained in future research.

References

  1. Pool SE, Krenning EP, Koning GA, van Eijck CH, Teunissen JJ, Kam B, Valkema R, Kwekkeboom DJ, de Jong M: Preclinical and clinical studies of peptide receptor radionuclide therapy. Semin Nucl Med 2010, 40: 209–218. 10.1053/j.semnuclmed.2009.12.001

    Article  PubMed  Google Scholar 

  2. Kwekkeboom DJ, Kam BL, van Essen M, Teunissen JJ, van Eijck CH, Valkema R, de Jong M, de Herder WW, Krenning EP: Somatostatin-receptor-based imaging and therapy of gastroenteropancreatic neuroendocrine tumors. Endocr Relat Cancer 2010, 17: R53-R73. 10.1677/ERC-09-0078

    Article  CAS  PubMed  Google Scholar 

  3. Kwekkeboom DJ, Bakker WH, Kooij PP, Konijnenberg MW, Srinivasan A, Erion JL, Schmidt MA, Bugaj JL, de Jong M, Krenning EP: [177Lu-DOTAOTyr3]octreotate: comparison with [111In-DTPAo]octreotide in patients. Eur J Nucl Med 2001, 28: 1319–1325. 10.1007/s002590100574

    Article  CAS  PubMed  Google Scholar 

  4. Sward C, Bernhardt P, Ahlman H, Wangberg B, Forssell-Aronsson E, Larsson M, Svensson J, Rossi-Norrlund R, Kolby L: [177Lu-DOTA 0-Tyr 3]-octreotate treatment in patients with disseminated gastroenteropancreatic neuroendocrine tumors: the value of measuring absorbed dose to the kidney. World Journal of Surgery 2010, 34: 1368–1372. 10.1007/s00268-009-0387-6

    Article  PubMed  Google Scholar 

  5. Melis M, Krenning EP, Bernard BF, Barone R, Visser TJ, de Jong M: Localisation and mechanism of renal retention of radiolabelled somatostatin analogues. Eur J Nucl Med Mol Imaging 2005, 32: 1136–1143. 10.1007/s00259-005-1793-0

    Article  CAS  PubMed  Google Scholar 

  6. Forrer F, Valkema R, Kwekkeboom DJ, de Jong M, Krenning EP: Neuroendocrine tumors. Peptide receptor radionuclide therapy. Best Pract Res Clin Endocrinol Metab 2007, 21: 111–129. 10.1016/j.beem.2007.01.007

    Article  CAS  PubMed  Google Scholar 

  7. Rolleman EJ, Kooij PP, de Herder WW, Valkema R, Krenning EP, de Jong M: Somatostatin receptor subtype 2-mediated uptake of radiolabelled somatostatin analogues in the human kidney. Eur J Nucl Med Mol Imaging 2007, 34: 1854–1860. 10.1007/s00259-007-0457-7

    Article  CAS  PubMed  Google Scholar 

  8. Moncayo R: Cubilin and megalin in radiation-induced renal injury with labelled somatostatin analogues: are we just dealing with the kidney? Eur J Nucl Med Mol Imaging 2005, 32: 1131–1135. 10.1007/s00259-005-1885-x

    Article  CAS  PubMed  Google Scholar 

  9. Rolleman EJ, Krenning EP, Van Gameren A, Bernard BF, De Jong M: Uptake of [111In-DTPA0]octreotide in the rat kidney is inhibited by colchicine and not by fructose. J Nucl Med 2004, 45: 709–713.

    CAS  PubMed  Google Scholar 

  10. Vegt E, de Jong M, Wetzels JF, Masereeuw R, Melis M, Oyen WJ, Gotthardt M, Boerman OC: Renal toxicity of radiolabeled peptides and antibody fragments: mechanisms, impact on radionuclide therapy, and strategies for prevention. J Nucl Med 2010, 51: 1049–1058. 10.2967/jnumed.110.075101

    Article  CAS  PubMed  Google Scholar 

  11. Verroust PJ, Christensen EI: Megalin and cubilin–the story of two multipurpose receptors unfolds. Nephrol Dial Transplant 2002, 17: 1867–1871. 10.1093/ndt/17.11.1867

    Article  CAS  PubMed  Google Scholar 

  12. Vegt E, Melis M, Eek A, de Visser M, Brom M, Oyen WJ, Gotthardt M, de Jong M, Boerman OC: Renal uptake of different radiolabelled peptides is mediated by megalin: SPECT and biodistribution studies in megalin-deficient mice. Eur J Nucl Med Mol Imaging 2011, 38: 623–632. 10.1007/s00259-010-1685-9

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  13. Rolleman EJ, Melis M, Valkema R, Boerman OC, Krenning EP, de Jong M: Kidney protection during peptide receptor radionuclide therapy with somatostatin analogues. Eur J Nucl Med Mol Imaging 2010, 37: 1018–1031. 10.1007/s00259-009-1282-y

    Article  PubMed  Google Scholar 

  14. Barone R, Borson-Chazot F, Valkema R, Walrand S, Chauvin F, Gogou L, Kvols LK, Krenning EP, Jamar F, Pauwels S: Patient-specific dosimetry in predicting renal toxicity with (90)Y-DOTATOC: relevance of kidney volume and dose rate in finding a dose-effect relationship. J Nucl Med 2005,46(Suppl 1):99S-106S.

    CAS  PubMed  Google Scholar 

  15. Cremonesi M, Ferrari ME, Bodei L, Bartolomei M, Chinol M, Mei R, Daou B, Tosi G, Paganelli G: Dosimetry in patients undergoing Lu-177-DOTATATE therapy with indications for Y-90-DOTATATE. Eur J Nucl Med Mol I 2006, 33: S102-S102.

    Article  Google Scholar 

  16. Fleming JS: A technique for the absolute measurement of activity using a gamma camera and computer. Phys Med Biol 1979, 24: 176–180. 10.1088/0031-9155/24/1/017

    Article  CAS  PubMed  Google Scholar 

  17. Wehrmann C, Senftleben S, Zachert C, Muller D, Baum RP: Results of individual patient dosimetry in peptide receptor radionuclide therapy with 177Lu DOTA-TATE and 177Lu DOTA-NOC. Cancer Biother Radiopharm 2007, 22: 406–416. 10.1089/cbr.2006.325

    Article  CAS  PubMed  Google Scholar 

  18. Valkema R, Pauwels SA, Kvols LK, Kwekkeboom DJ, Jamar F, de Jong M, Barone R, Walrand S, Kooij PP, Bakker WH, Lasher J, Krenning EP: Long-term follow-up of renal function after peptide receptor radiation therapy with (90)Y-DOTA(0), Tyr(3)-octreotide and (177)Lu-DOTA(0), Tyr(3)-octreotate. J Nucl Med 2005,46(Suppl 1):83S-91S.

    CAS  PubMed  Google Scholar 

  19. Sandstrom M, Garske U, Granberg D, Sundin A, Lundqvist H: Individualized dosimetry in patients undergoing therapy with (177)Lu-DOTA-D-Phe (1)-Tyr (3)-octreotate. Eur J Nucl Med Mol Imaging 2010, 37: 212–225. 10.1007/s00259-009-1216-8

    Article  PubMed  Google Scholar 

  20. Garkavij M, Nickel M, Sjogreen-Gleisner K, Ljungberg M, Ohlsson T, Wingardh K, Strand SE, Tennvall J: 177Lu-[DOTA0, Tyr3] octreotate therapy in patients with disseminated neuroendocrine tumors: analysis of dosimetry with impact on future therapeutic strategy. Cancer 2010, 116: 1084–1092. 10.1002/cncr.24796

    Article  CAS  PubMed  Google Scholar 

  21. Norrgren K, Svegborn SL, Areberg J, Mattsson S: Accuracy of the quantification of organ activity from planar gamma camera images. Cancer Biother Radiopharm 2003, 18: 125–131. 10.1089/108497803321269403

    Article  PubMed  Google Scholar 

  22. Buijs WC, Siegel JA, Boerman OC, Corstens FH: Absolute organ activity estimated by five different methods of background correction. J Nucl Med 1998, 39: 2167–2172.

    CAS  PubMed  Google Scholar 

  23. ICRP89: Basic anatomical and physiological data for use in radiological protection: reference values. ICRP Publication 89 Ann ICRP 2002, 32: 1–277.

    Google Scholar 

  24. ICRP38: Radionuclide transformations—energy and intensity of emissions. ICRP Publication 38 Ann ICRP 1983, 11–13: 738–761.

    Google Scholar 

  25. ICRP107: Nuclear decay data for dosimetric calculations. ICRP Publication 2008, 38: 1–96.

    Google Scholar 

  26. Nuclear Decay Data in the MIRD Format Database. . Accessed March 14, 2012 http://nucleus.iaea.org/sso/NUCLEUS.html?exturl=http://www-nds.iaea.org/mird/

  27. Enger SA, Hartman T, Carlsson J, Lundqvist H: Cross-fire doses from beta-emitting radionuclides in targeted radiotherapy. A theoretical study based on experimentally measured tumor characteristics. Phys Med Biol 2008, 53: 1909–1920. 10.1088/0031-9155/53/7/007

    Article  CAS  PubMed  Google Scholar 

  28. Amato E, Lizio D, Baldari S: Absorbed fractions for electrons in ellipsoidal volumes. Phys Med Biol 2011, 56: 357–365. 10.1088/0031-9155/56/2/005

    Article  CAS  PubMed  Google Scholar 

  29. Amato E, Lizio D, Baldari S: Absorbed fractions for photons in ellipsoidal volumes. Phys Med Biol 2009, 54: N479-N487. 10.1088/0031-9155/54/20/N02

    Article  CAS  PubMed  Google Scholar 

  30. Snyder WS, Ford MR, Warner GG: MIRD pamphlet no. 5, revised: Estimates of specific absorbed fractions for photon sources uniformly distributed in various organs of a heterogeneous phantom. New York: Society of Nuclear Medicine; 1978.

    Google Scholar 

  31. Stabin MG, Siegel JA: Physical models and dose factors for use in internal dose assessment. Health physics 2003, 85: 294–310. 10.1097/00004032-200309000-00006

    Article  CAS  PubMed  Google Scholar 

  32. Bodei L, Cremonesi M, Ferrari M, Pacifici M, Grana CM, Bartolomei M, Baio SM, Sansovini M, Paganelli G: Long-term evaluation of renal toxicity after peptide receptor radionuclide therapy with 90Y-DOTATOC and 177Lu-DOTATATE: the role of associated risk factors. Eur J Nucl Med Mol Imaging 2008, 35: 1847–1856. 10.1007/s00259-008-0778-1

    Article  CAS  PubMed  Google Scholar 

  33. Forssell-Aronsson E, Lanhede B, Fjälling M, Wängberg B, Tisell LE, Ahlman H, Mattsson S: Pharmacokinetics and dosimetry of 111In-DTPA-D-Phe1-octreotide in patients with neuroendocrine tumors. In Sixth International Radiopharmaceutical Dosimetry Symposium: May 7–10, 1996; Gatlinburg, Tennessee. volume 2. Edited by: Schlafke-Stelson A, Watson EE. Tennessee: ORISE 99–0164; 1999:643–654.

    Google Scholar 

  34. Jonsson L, Ljungberg M, Strand SE: Evaluation of accuracy in activity calculations for the conjugate view method from Monte Carlo simulated scintillation camera images using experimental data in an anthropomorphic phantom. J Nucl Med 2005, 46: 1679–1686.

    PubMed  Google Scholar 

Download references

Acknowledgments

This study was supported by the grants from the Swedish Research Council, the Swedish Cancer Society, BioCARE (a National Strategic Research Program at University of Gothenburg), the King Gustav V Jubilee Clinic Cancer Research Foundation, and the Assar Gabrielsson Cancer Research Foundation. The work was performed within the EC COST Action BM0607.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Maria Larsson.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

BW, HA, EFA, PB, and JS planned the therapeutic study, and ML, PB and EFA planned the dosimetric study. ML performed the calibrations, image analyses, and the first compilation of the data. ML and EFA made the analysis and interpretation of the data. All authors contributed to the scientific and intellectual discussion. ML drafted the manuscript. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Larsson, M., Bernhardt, P., Svensson, J.B. et al. Estimation of absorbed dose to the kidneys in patients after treatment with 177Lu-octreotate: comparison between methods based on planar scintigraphy. EJNMMI Res 2, 49 (2012). https://doi.org/10.1186/2191-219X-2-49

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/2191-219X-2-49

Keywords